Order No.: 10079352/02 Operating Manual ALTAIR PRO Single Gas Detector

Page 2

GBDESCRIPTION MSA 10 ALTAIR PRO - Operating Manual 2.2. On-Screen Indicators Fig. 2 Display 1 Self-test in progress 2 Bump check indicator

Page 3 - Declaration of Conformity

GBMSA DESCRIPTION ALTAIR PRO - Operating Manual 11 Battery Warning A battery warning indicates that a nominal two days of operation remain befo

Page 4

GBDESCRIPTION MSA 12 ALTAIR PRO - Operating Manual Confidence indicator The Confidence indicator flashes once every 60 seconds to notify the user

Page 5

GBMSA DESCRIPTION ALTAIR PRO - Operating Manual 13 The backlight turns ON for 20 seconds during an alarm condition. If the gas concentration r

Page 6 - 1. Safety Regulations

GBDESCRIPTION MSA 14 ALTAIR PRO - Operating Manual 2.5. Duration of Use of the Instrument Normal Operation The instrument is ready for operation

Page 7

GBMSA OPERATION ALTAIR PRO - Operating Manual 15 3. Operation 3.1. Changing Alarm Setpoints Alarm setpoints can only be changed manually dur

Page 8 - 2. Description

GBOPERATION MSA 16 ALTAIR PRO - Operating Manual High Alarm Setpoint for three seconds, "HI" and "ALARM" icons turn ON. (8) T

Page 9

GBMSA OPERATION ALTAIR PRO - Operating Manual 17 Fig. 3 ALTAIR PRO Setup

Page 10 - 2.2. On-Screen Indicators

GBOPERATION MSA 18 ALTAIR PRO - Operating Manual 3.2. Turning ON the Instrument Before the instrument can be used for monitoring possible toxic

Page 11

GBMSA OPERATION ALTAIR PRO - Operating Manual 19 When performing a fresh air setup, the instrument must be in a known clean air. The fresh ai

Page 12 - 2.3. Monitoring Toxic Gases

MSA AUER GmbH D-12059 Berlin Thiemannstrasse 1 Germany © MSA AUER GmbH. All rights reserved

Page 13

GBOPERATION MSA 20 ALTAIR PRO - Operating Manual Fig. 4 Turning ON the instrument

Page 14 - ERR"

GBMSA OPERATION ALTAIR PRO - Operating Manual 21 3.3. Toxic Gas Measurements The ALTAIR PRO can be purchased to detect the following gases in

Page 15 - 3. Operation

GBOPERATION MSA 22 ALTAIR PRO - Operating Manual High Alarm If the gas concentration reaches or exceeds the high alarm setpoint the instrument wil

Page 16

GBMSA OPERATION ALTAIR PRO - Operating Manual 23 3.4. Oxygen Measurements The ALTAIR PRO Detector can be purchased to measure the concentratio

Page 17 - Fig. 3 ALTAIR PRO Setup

GBOPERATION MSA 24 ALTAIR PRO - Operating Manual 3.5. Displaying the Instrument Data - excluding the Steel Mill Version A single, quick press on

Page 18

GBMSA OPERATION ALTAIR PRO - Operating Manual 25 6) IR mode When instrument displays "IR?", press the button to enter IR mode. I

Page 19 - FAS/ERR"

GBOPERATION MSA 26 ALTAIR PRO - Operating Manual 11) Time Weighted Average ("TWA") The TWA reading displays what the instrument has c

Page 20 - OPERATION MSA

GBMSA OPERATION ALTAIR PRO - Operating Manual 27 Fig. 5 Displaying instrument data (does not apply to the Steel Mill Version)

Page 21 - 3.3. Toxic Gas Measurements

GBOPERATION MSA 28 ALTAIR PRO - Operating Manual 3.8. Data Logging Session Log The ALTAIR PRO Detector has the capability of recording 50 of the

Page 22 - HI" icon (if the high

GBMSA OPERATION ALTAIR PRO - Operating Manual 29 3.9. Function Tests on the Instrument Checking the Confidence indicator The alarm LEDs and t

Page 23 - 3.4. Oxygen Measurements

GBMSA DECLARATION OF CONFORMITY ALTAIR PRO - Operating Manual 3 Declaration of Conformity MANUFACTURED BY: Mine Safety

Page 24

GBOPERATION MSA 30 ALTAIR PRO - Operating Manual If the "" does not appear and "ERR" is displayed, check:  whether the sens

Page 25

GBMSA OPERATION ALTAIR PRO - Operating Manual 31 3.10. Calibrating the Instrument The ALTAIR PRO must be calibrated if it does not pass a bump

Page 26 - CLR" is displayed

GBOPERATION MSA 32 ALTAIR PRO - Operating Manual Carry out the calibration as follows (also see Fig. 7and Fig. 8): (1) Press the button.  "

Page 27 - MSA OPERATION

GBMSA OPERATION ALTAIR PRO - Operating Manual 33 (7) If calibration fails, check:  whether the correct test gas was used,  whether the test

Page 28 - CONNECT"

GBOPERATION MSA 34 ALTAIR PRO - Operating Manual Fig. 7 Calibration - Toxic gases

Page 29 - OK" is displayed

GBMSA OPERATION ALTAIR PRO - Operating Manual 35 Calibration - Oxygen Attention! The instrument must be calibrated if, during operation, a val

Page 30

GBOPERATION MSA 36 ALTAIR PRO - Operating Manual Fig. 8 Calibration – Oxygen

Page 31

GBMSA MAINTENANCE ALTAIR PRO - Operating Manual 37 4. Maintenance If irregularities occur during operation, use the displayed error codes to de

Page 32

GBMAINTENANCE MSA 38 ALTAIR PRO - Operating Manual 4.2. Replacing the Battery Fig. 9 Sensor and Battery Replacement 1 Sensor 2 Battery

Page 33 - " is displayed

GBMSA MAINTENANCE ALTAIR PRO - Operating Manual 39 4.3. Replacing the Sensor Caution! Before handling the PC board, ensure you are properly

Page 34

GBCONTENTS MSA 4 ALTAIR PRO Single Gas Detector Contents 1. Safety Regulations ...

Page 35

GBTECHNICAL SPECIFICATIONS/CERTIFICATIONS MSA 40 ALTAIR PRO - Operating Manual 5. Technical Specifications/Certifications 5.1. Technical Specifi

Page 36 - Fig. 8 Calibration – Oxygen

GBMSA TECHNICAL SPECIFICATIONS/CERTIFICATIONS ALTAIR PRO - Operating Manual 41 Temperature range Operation : Storage : Vibration alarm : -

Page 37 - 4. Maintenance

GBTECHNICAL SPECIFICATIONS/CERTIFICATIONS MSA 42 ALTAIR PRO - Operating Manual 5.2. Certifications European Community The product ALTAIR PRO com

Page 38 - 4.2. Replacing the Battery

GBMSA TECHNICAL SPECIFICATIONS/CERTIFICATIONS ALTAIR PRO - Operating Manual 43 5.3. Test Gases Table Instrument version Test gas - Self-test

Page 39 - 4.3. Replacing the Sensor

GBTECHNICAL SPECIFICATIONS/CERTIFICATIONS MSA 44 ALTAIR PRO - Operating Manual Sensor SO2 (Sulfur Dioxide) Range 0-20.0 ppm Resolution 0.1 pp

Page 40 - Link™ Software

GBMSA TECHNICAL SPECIFICATIONS/CERTIFICATIONS ALTAIR PRO - Operating Manual 45 5.6. Data Logging Specifications Session datalog Number of shar

Page 41

GBORDER INFORMATION MSA 46 ALTAIR PRO - Operating Manual 6. Order Information Description Part No. Test gas cylinder, 60 ppm CO (factory setting

Page 42 - 5.2. Certifications

GBMSA ORDER INFORMATION ALTAIR PRO - Operating Manual 47 Description Part No. Sensor, O2 10046946 Sensor, CO, Steel Mill, H2 resistant 10074

Page 43 - 5.3. Test Gases Table

MSA in Europe[ www.msa-europe.com & www.msa-gasdetection.com ]Northern Europe Southern Europe Eastern Europe Central EuropeNetherlandsMSA Nederlan

Page 44 - only: -20 °C to 0 °C)

GBMSA CONTENTS ALTAIR PRO - Operating Manual 5 5. Technical Specifications/Certifications...

Page 45

GBSAFETY REGULATIONS MSA 6 ALTAIR PRO - Operating Manual 1. Safety Regulations 1.1. Correct Use The ALTAIR PRO is a portable instrument for mon

Page 46 - 6. Order Information

GBMSA SAFETY REGULATIONS ALTAIR PRO - Operating Manual 7 1.3. Safety and Precautionary Measures to be Adopted Attention! The following safet

Page 47

GBDESCRIPTION MSA 8 ALTAIR PRO - Operating Manual 2. Description 2.1. Instrument Overview Fig. 1 Instrument view 1 Alarm LED 5 Button 2 I

Page 48 - MSA in Europe

GBMSA DESCRIPTION ALTAIR PRO - Operating Manual 9 The response levels for the individual gases are set at the factory and cannot be changed dur