ALTAIR PRO - Operating Manual 19
When performing a fresh air setup, the instrument must be in a known
clean air. The fresh air setup has limits.
If a hazardous level of gas is present, the instrument displays "
FAS/ERR".
Press the button to acknowledge the error and perform an instrument
calibration.
(4) If Fresh Air Setup is desired, immediately press the button.
"
HOURGLASS", "SET" and "FAS" display.
(5) If Fresh Air Setup is not desired, do not press the button.
The instrument continues the ON sequence.
(6) If the instrument was configured for Oxygen, it displays
Oxygen reading,
% icon and
battery indicator.
(7) If the instrument was configured for a toxic gas, it displays
gas reading,
ppm icon and
battery indicator.