OPERATION MSA
18 ALTAIR PRO - Operating Manual
3.2. Turning ON the Instrument
Before the instrument can be used for monitoring possible toxic gases in
the atmosphere, it must be commissioned ( Fig. 4).
(1) Press and hold the button for 3 seconds.
"
ON" displays during this time period.
The Steel Mill version cannot be turned OFF without removing the battery
or using MSA FiveStar
®
Link
Software.
An LCD functional test activates all display indicators.
Acoustic alarm sounds, alarm LEDs light up and vibration alarm is
triggered.
The software version is displayed for approx. 3 seconds.
The gas type is displayed for approx. 3 seconds more
(O2, H
2
S, CO, SO
2
, NO
2
, etc).
LO and ALARM as well as HI and ALARM are displayed alternately for
approx. 3 seconds.
STL and ALARM followed by the STEL alarm setpoint are displayed for
approx. 3 seconds.
TWA and ALARM followed by the TWA alarm setpoint are displayed for
approx. 3 seconds.
During the display of
LO,HI,STL,TWA and ALARM the
corresponding setpoints can still be changed ( Fig. 4).
(2) If Calibration Due (CAL DUE) is enabled via MSA FiveStar
®
Link
Software
(default is
OFF):
"
CAL" and the hourglass display.
If calibration is due, "
DUE" and the hourglass display for three seconds.
The instrument beeps and flashes "CAL" "DUE" every minute until
instrument calibration is performed.
If calibration is not due, the hourglass, number of days to calibration, and
"
DAYS" display.
(3) User is prompted for a Fresh Air Setup (FAS).
"
SET", "?" and "FAS" display.