OPERATION MSA
32 ALTAIR PRO - Operating Manual
Carry out the calibration as follows (also see Fig. 7and Fig. 8):
(1) Press the button.
"
TEST" "GAS" "?" is displayed.
(2) Press and hold the button for 3 seconds.
"
TEST" "CAL" is displayed.
After 3 seconds, "
FAS" "?" is displayed which prompts the user to do a
fresh air equalisation.
(3) During the "
FAS" "?" display, press the button to do a fresh air equalisation.
If no fresh air equalisation is carried out, the instrument returns to
normal operating mode.
During the equalisation, the hour glass and "FAS" are displayed.
If the equalisation fails ("
ERR" display), the instrument returns to normal
operating mode after 5 s.
(4) If the equalisation is successful ("
OK" display), press the button and begin
the calibration.
"
CAL" "?" is displayed.
(5) During the "
CAL" "?" display, press the button and start the calibration
mode.
The value for the test gas in ppm is displayed.
(6) To change the value, press and hold the button, until the desired value is
set.
The new value for the test gas in ppm is displayed.
The button can be held down to change in larger increments.
After 3 seconds, the instrument returns to the calibration mode.
(6) Feed test gas into the instrument (possible test gases Section 5.3).
Attention!
The concentration of the test gas used must match the stipulated
value in the "Test gases" table in Section 5.3 for the relevant
instrument.
Otherwise, the calibration is not carried out correctly. If there is a fault,
it can lead to serious damage to health or even to death.
The gas value, hour glass and "CAL" are displayed alternately.
If calibration is successful, after approx. 90 s "
OK" is displayed and the
instrument returns to normal operating mode after 5 seconds.
If calibration fails, "
ERR" is displayed and the instrument returns to
normal operating mode after 5 seconds.
The current values were not changed.