WARNING
If the TOXIC GAS alarm condition is reached while using the
instrument as a personal or area monitor, leave contaminated
area immediately; the ambient gas concentration has reached
a preset alarm level. If using the instrument as an inspection
device, do not enter area without proper protection. Failure
to follow this warning will cause over-exposure to toxic gases
which can result in serious personal injury or death.
Calibration Check
This calibration check is very simple
and should only take one to five
minutes, depending on the number
and type of gases your FiveStar
Alarm is equipped to sense. Turn the
FiveStar Passport FiveStar Alarm ON
in clean fresh air, and verify that the
readings indicate no gas present.
1. Attach calibration cap to the
FiveStar Alarm, orienting the
inlet fitting to point toward the
battery pack:
2. Attach the calibration adapter
to the calibration cap.
3. Attach the regulator (supplied
with the calibration kit) to
the cylinder.
4. Connect the black tubing
supplied with the calibration kit
to the regulator.
5. Open the valve on the regulator,
and connect the other end of the
tubing to the inlet fitting.
The flow rate of the regulator is 0.25
lpm. Note the readings on the
FiveStar display; they should be within
the limits stated on the calibration
cylinder or limits determined by your
company. (If necessary, change
Figure 2-28.
Metal Calibration Cap Installation
(Early Versions Only)
Figure 2-29.
Snap Calibration Cap Installation
Instruction Manual Chapter 2, Using the FiveStar Alarm
FiveStar Alarm 2-15