MSA FiveStar Instruction Manual Page 43

  • Download
  • Add to my manuals
  • Print
  • Page
    / 71
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 42
"
WARNING
Certain gases may react with the Aspirator Assembly; this
may cause lower than actual readings. If these gases are
suspected, re-verify readings in the diffusion mode by removing
the Aspirator Assembly or by using Pump Module (P/N 710790).
Attaching the Aspirator Assembly
1. Remove the two sensor lid screws located closest to the
display face.
Store screws inside the Aspirator Assembly, on the white
manifold containing two recesses on each edge to
securely hold the screws.
2. Once screws are removed, align the Aspirator Assembly tabs
to sit inside the main body recesses by tilting the Aspirator
Assembly gently onto the back of the FiveStar unit.
Ensure tabs engage notches; do not force connection as
damage can occur.
3. Gently rotate the Aspirator Assembly until the
manifold contacts the FiveStar unit sensor cover.
Thumb screws align with holes left by the screws removed
in step 1.
4. Finger-tighten the screws to secure the Aspirator Assembly to
the FiveStar unit.
Do not use tools or force the screws as damage can occur.
5. Attach sample lines (up to 25 feet long) and attach the probe.
Operation and Use
1. Block sample line or probe inlet and squeeze the aspirator bulb.
With long sample line lengths (up to 25 feet), the bulb may
slowly inflate as a vacuum is drawn in the sample line;
squeeze the bulb one more time if necessary.
The bulb should now remain deflated; if bulb inflates,
check for leaks and tighten as necessary.
2. Place the end of the probe at the point the sample is to be
taken.
3. Squeeze the bulb 10 times, allowing it to fully inflate
between squeezes.
Note the readings.
Instruction Manual Chapter 2, Using the FiveStar Alarm
FiveStar Alarm 2-27
Page view 42
1 2 ... 38 39 40 41 42 43 44 45 46 47 48 ... 70 71

Comments to this Manuals

No comments